PERFORMING A CONTROL SOLUTION TEST
Complete steps 1-6, first using Level 1 control solution, again using Level 2 control solution.
1
2
Step 1: To turn meter on, insert test strip as shown. Step 2: Press the Back or Forward button one time to enter the control solution mode. The control solution bottle will appear in upper right portion of the screen. Step 3: a. Gently mix Control Solution by inverting three times b. Discard first drop c. Apply second drop to cap
3
4
Step 4: Bring meter and strip to drop. Test strip will draw up the solution. Step 5: Compare result to the range printed on test strip bottle.
5
Date Opened (Fecha de Apertura)
(Use within 90 days of first opening) (Usar dentro de los 90 dias a partir de la fecha en que fue abierto)
446 11182A 2010-01
CODE
Control Solution Range (mg/dL) (Rango de la solución de control)
Level 1: Level 2:
80-120 200-300
Step 6: Push strip release button forward and discard strip.